Jasco V-670 UV-VIS-NIR Molecular Absorption Spectrometer
Capabilities:
- Wavelength Range: 190–3200 nm
- Radiation Sources: Deuterium lamp (190–350 nm); Halogen lamp (350–3200 nm)
- Detector: PMT and PbS photoconductive cell
- Cells: Silica, Optical Glass, and Plastic
Information on Silica Cells:
Type | Wavelength Range (nm) | Vol. (mL) |
1-l-10 | 220-3800 | 3.5 |
18/9-Q-10 | 170-2700 | 0.580 |
- Information on cleaning cells
- Peltier temperature control of cells
- Must turn on circulating water bath when Peltier cooler used
- Nitrogen gas flow must be used in cell holder when using temperatures at or below 15 °C
- From Model V-630/650/660/670 Spectrometer Hardware/Function Manual
- Instrument Specifications
- Optical Configuration
Software:
- Spectra Manager Ver. 2.08.04 {Build 1}
Other Resources:
- Simple Operating Instructions
- Peltier Temperature Control System
- Heat Exchange System: Koolance EX2-1055
.